DLS measures particle size distributions in solutions and suspensions.

Nanoparticles, biopolymers, environmental monitoring, macromolecules

Measurements

The DynaPro NanoStar™ from Wyatt Technology is Dynamic Light Scattering (DLS) instrument that is used for the analysis of protein solutions, aggregates, buffers, nanoparticles, polymers, or other products in solution. In a DLS measurement, the time-dependent fluctuations in the scattered light are measured by a fast photon counter. The fluctuations are directly related to the rate of diffusion of the molecule through the solvent. Therefore, the fluctuations can be analyzed to determine a hydrodynamic radius for the sample (0,5 nm to 1000 nm). Molar mass range: 300 Da to 106 Da. Minimum concentration = 0,1 mg/ml.
 

Equipment
Instrument: DynaPro NanoStar™ from Wyatt Technology
Temperature: 22°C-150°C
Laser wavelength: 662 nm
Laser power: 100 mW

Cuvette: 1 µl (standard quartz-cuvette). Disposable cuvettes available (3 µl and 50 µl)
 

Interested?

Get trained or get more info about sending samples for analysis by filling out the applicable form below. Lab staff will follow up with more details. 